Validation of an analytical method to quantify serum electrolytes by atomic absorption spectroscopy
ISSN 0188-6266
doi: 10.15174/au.2015.747
Validation of an analytical method to quantify serum
electrolytes by atomic absorption spectroscopy
Validación de un método analítico para cuantificar electrolitos séricos
por espectroscopia de absorción atómica
Alejandro Monserrat García-Alegría*, Agustín Gómez-Álvarez**, Leticia García-Rico***, Mercedes Serna-Félix****
ABSTRACT
The objective of this research was to develop the procedure to validate an analytical method
through the technique of spectroscopy by atomic absorption to quantify serum electrolytes.
Quality parameters were used such as: accuracy, precision, repeatability, linearity, limit
of detection, and limit of quantification. First some work was done in optimization of
the analytical method in ideal conditions. Subsequently, some work was done with the
validation of the method using a certified reference material National Institute Standard
Technology (NIST) 2670a. Finally, the methodology was assessed in the determination of
metals in a sample of human blood serum. The quality criteria evaluated showed consistent and
acceptable values depending on the statistical test used for each parameter with a significance
level of 5% (α = 0.05). The results obtained in this research can contribute to the beginning
of the establishment of an official standard aimed at analysis of electrolytes (Na, K, Ca, Mg) in
human blood serum so that can be used as reference by clinical laboratories, public health
institutions, as well as in institutions of higher education and research centers.
RESUMEN
El objetivo fue validar un método analítico mediante la técnica de espectroscopía de absorción
atómica para cuantificar electrolitos séricos, utilizando parámetros analíticos como exactitud,
precisión, repetitibilidad, linealidad, límite de detección y límite de cuantificación. Primeramente se trabajó en la optimización del método analítico en condiciones ideales. Posteriormente
en la validación del método utilizando un material de referencia certificado por el National
Institute Standard Technology (NIST) 2670a. Para finalizar, se evaluó la metodología en la determinación de metales en una muestra de suero sanguíneo humano. Los criterios de calidad
evaluados mostraron valores consistentes y aceptables en función del estadístico de prueba
empleado para cada parámetro con un nivel de significancia del 5% (α = 0.05). Los resultados obtenidos pueden contribuir al inicio del establecimiento de una norma oficial mexicana
encaminada al análisis de electrolitos (Na, K, Ca, Mg) en suero sanguíneo humano para que
pueda ser utilizada como referencia por laboratorios de análisis clínicos, instituciones de
salud pública, así como en instituciones de educación superior y en centros de investigación.
Recibido: 7 de abril de 2015
Aceptado: 8 de mayo de 2015
Keywords:
Validation; optimization; serum electrolytes; atomic absorption spectroscopy.
Palabras clave:
Validación; optimización; electrolitos séricos; espectroscopia de absorción atómica.
Cómo citar:
García-Alegría, A. M., Gómez-Álvarez, A., GarcíaRico, L. & Serna-Félix, M. (2015). Validation of an
analytical method to quantify serum electrolytes by
atomic absorption spectroscopy. Acta Universitaria, 25(3), 3-12. doi: 10.15174/au.2015.747
INTRODUCTION
The fundamental purpose of a clinical laboratory is to generate information
that enables the diagnosis of diseases, as well as their prevention and treatment. In order to do so, it is necessary to use reliable, accurate and appropriate analytical methods for this purpose. It is required to optimize the
experimental conditions, through the standardization of the analytical method, which determines the efficiency of the test from the different conditions
in which the purpose is to quantify an analyte and then proceed to validation
(Centro Nacional de Metrología-Entidad Mexicana de Acreditación [CENAMEMA]), 2008a; Coy, 1999; Díaz-Romero, Henríquez-Sánchez, López-Blanco,
Rodríguez-Rodríguez & Serra-Majem, 2002). Validation is confirmation that
through provision of objective evidence they have met the requirements of the
method for a specific application or intended use Intgernational Organization
for Standardization/International Electrotechnical Commission ([ISO/IEC]
* Chemical Department of Biological Sciences, University of Sonora. Luis Encinas and Rosales n/n. Colony Center, Hermosillo, Sonora, México, C.P. 83000. E-mail:
** Department of Chemical Engineering and Metallurgy, University of Sonora, Hermosillo, Sonora, México, C.P.: 83000. Phone: (662) 2592165. E-mail:
*** Research Center for Food and Development Civil Association (CA). Road La Victoria n/n. Colony Center, Hermosillo, Sonora, México, C.P. 83000. E-mail:
**** School of Nursing of the Mexican Institute of Social Security. Hermosillo, Sonora, México. Ortíz Mena and Justo Sierra n/n. Colony Modelo, Hermosillo, Sonora, México, C.P. 83000. E-mail:
Vol. 25 No. 3 Mayo-Junio 2015
3
ISSN 0188-6266
17025, 2005; NMX-Z-055-IMNC-1996; NMX-CH-152IMNC-2005). Between the analytes that are determined
in a clinical laboratory and that require the provision
of safe and reliable data are the serum electrolytes (Na,
K, Ca and Mg). These have important functions in the
maintenance of body homeostasis. When the levels
of these electrolytes increase or decrease problems
such as diarrhea, dehydration, tachycardia, kidney failure, seizures, tetany, infarcts, and others might arise
(Boetefür & Müller-Plathe, 1995; Spelch, Bousquet &
Nicolas, 1980). Normally these electrolytes are quantified through spectrophotometric methods that have a
lower sensitivity and low specificity, as well as a higher
limit of detection and quantification and increased risk
of interference compared to atomic absorption spectroscopy (Baruthio & Pierre, 1993; Olmedo et al., 2010).
Additionally, there are differences in the concentrations of analyzed metals and reported inter-laboratory
(Namkoong, Hong, Kim & Park, 2013). Due to the reasons mentioned above and to the need for a reference
method to quantify metals, the aim of the present work is
to validate the methodology to quantify serum electrolytes
(Na, K, Ca and Mg) by atomic absorption spectroscopy.
MATERIALS AND METHODS
Reagents and materials
All the solutions were prepared with deionized water and chemical reagents used (HNO3 and H2O2) are
high-purity (Merck).
Standard solutions of each metal. They are prepared
each time the determination by atomic absorption is
made, from a stock solution of 1000 mg/L analytical
grade (Na cas # 7440-23-5, K cas # 7440-09-7, Ca cas
# 7440-70-2 and Mg cas # 7439-95-4 AccuStandard)
in 0.2% of HNO3.
Chemical modifiers such as LiCl and LaCL3 to 0.2%
are used, to avoid ionic interference in Na and K, as well
as chemical interference in Ca and Mg, respectively.
containers were washed with liquid detergent rinsed
with plenty of deionized water, subsequently they were
subjected to a solution of HNO3 to 30% for three days
to finally be rin (...truncated)