Retrieval of a partially degloved stent strut during percutaneous coronary intervention.
Chonnam Med J. 2012 Aug; 48(2): 130–132.
Published online 2012 Aug 24. doi: 10.4068/cmj.2012.48.2.130
PMCID: PMC3434794
PMID: 22977756
Retrieval of a Partially Degloved Stent Strut During Percutaneous Coronary Intervention
Woohyeun Kim, Yong Hyun Kim, Sang Yup Lim, Seong Hwan Kim, Jeong Chun Ahn, and Woo Hyuk Song
Woohyeun Kim
Department of Internal Medicine, Korea University Ansan Hospital, Ansan, Korea.
Find articles by Woohyeun Kim
Yong Hyun Kim
Department of Internal Medicine, Korea University Ansan Hospital, Ansan, Korea.
Find articles by Yong Hyun Kim
Sang Yup Lim
Department of Internal Medicine, Korea University Ansan Hospital, Ansan, Korea.
Find articles by Sang Yup Lim
Seong Hwan Kim
Department of Internal Medicine, Korea University Ansan Hospital, Ansan, Korea.
Find articles by Seong Hwan Kim
Jeong Chun Ahn
Department of Internal Medicine, Korea University Ansan Hospital, Ansan, Korea.
Find articles by Jeong Chun Ahn
Woo Hyuk Song
Department of Internal Medicine, Korea University Ansan Hospital, Ansan, Korea.
Find articles by Woo Hyuk Song
Author information Article notes Copyright and License information Disclaimer
Department of Internal Medicine, Korea University Ansan Hospital, Ansan, Korea.
Corresponding author.
Corresponding Author: Sang Yup Lim. Department of Internal Medicine, Korea University Ansan Hospital, 516 Gojan-dong, Danwon-gu, Ansan 425-707, Korea. TEL: +82-31-412-6755, FAX: +82-31-412-5594, ten.liamnah@revolnv
Received 2012 May 22; Revised 2012 Jul 18; Accepted 2012 Jul 23.
Copyright © Chonnam Medical Journal, 2012
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract
Although stent entrapment is a rare event during percutaneous coronary intervention, stent entrapment can cause stent breakage or loss, which results in fatal complications such as stent embolism or acute myocardial infarction. We report one case of stent entrapment that was successfully treated by a snare via a contralateral transfemoral approach.
Keywords: Stents, Angioplasty, Complications
INTRODUCTION
Complications during percutaneous coronary intervention (PCI) include coronary artery complications (perforation, distal embolization, side branch occlusion, and stent thrombosis) and vascular complications (puncture site hematoma, retroperitoneal bleeding, and atheroembolism). Among them, stent entrapment with resulting stent deployment failure is rare but can cause fatal conditions such as stent thrombosis or myocardial infarction.
Although many methods have been used for stent retrieval, there are few reports of the use of a snare via a contralateral femoral approach. Here we report one case of half-degloved stent entrapment that occurred during PCI and that was successfully treated by a snare via a contralateral approach.
CASE REPORT
An 80-year-old male was admitted owing to recurrent chest pain for elective PCI of a chronic total occlusion (CTO) in a left circumflex (LCX) lesion. Six months previously, the patient had undergone emergent PCI in the left anterior descending artery (LAD) owing to unstable angina. Follow-up coronary angiography revealed no significant in-stent restenosis in the LAD and CTO in the LCX (Fig. 1). We decided to revascularize the LCX lesion. A 7-Fr JL4 guiding catheter was engaged and predilation with a 2.0×20-mm balloon was done. After ballooning, we attempted to insert the 2.5×30-mm Resolute-Integrity (Medtronic, Minneapolis, MN, USA) stent into the LCX, but the stent failed to cross the CTO lesion in the LCX. We used the deep intubation technique, but the stent could not cross the lesion entirely because the lesion was severely calcified and angulated. While attempting to push the stent, we noticed that the stent was stuck within the lesion. After a struggle, the guiding catheter was taken out but we observed that the stent was half-degloved off the balloon (Fig. 2). Because of the risk of stent loss and embolization, we decided to pull out the catheter and stent strut by use of an Amplatz GooseNeck snare (EV3, Plymouth, MN, USA). We pushed down the guiding catheter and stent strut together to the iliac bifurcation level. The left femoral artery was punctured and we used a 6Fr Balkin sheath to grab the stent with the snare and both the stent strut and the balloon catheter were removed successfully by the snare (Fig. 3). We completed the procedure and decided on medical therapy instead of PCI for the LCX lesion. Two days later, the patient was discharged and there have been no adverse events during 3 months of clinical follow-up.
Open in a separate window
FIG. 1
Coronary angiography showed no significant in-stent restenosis in the left anterior descending artery (LAD) and chronic total occlusion in the left circumflex artery (LCX).
Open in a separate window
FIG. 2
During percutaneous coronary intervention, a half-degloved stent strut was observed in the LCX (arrow).
Open in a separate window
FIG. 3
A half-degloved stent strut and balloon catheter that were removed by use of a loop snare.
DISCUSSION
Entrapment of coronary angioplasty devices, especially the stent strut, is rare but is a serious complication of PCI for which cardiac surgery is sometimes required. Calcified, long lesions and angulated lesions may predispose to entrapment of the stent strut. The incidence of stent dislodgement or migration ranges from 1.4% to 8%.1-4 To prevent stent dislodgement or migration, predilatation with a balloon catheter or use of stents with a smaller profile and good tractability are recommended in case of severe angulated or calcified lesions, as in this patient.5
A variety of stent retrieval methods have been described in the literature, such as using a snare device, a multipurpose basket, a variety of forceps, an Angioguard (a distal protection device), a simple balloon and just crushing of the stent into the vessel wall by a balloon.6-10
However, using a balloon or a snare to rescue an entrapped stent in a half-degloved state may contribute to stent loss; to prevent stent loss, it would be safer to approach both sides by inserting the snare via another vascular route.
In this case, we pushed down the stent strut and guiding catheter together to the iliac bifurcation level. After that, the contralateral femoral artery was punctured and both stent strut and balloon catheter were removed successfully by a snare with the aid of a 6Fr Balkin sheath. We think that this technique is safer than other methods in this situation to prevent stent loss and embolization.
However, to avoid serious complications, interventional cardiologists should keep in mind the various complications during PCI (...truncated)