Quantitative comparison of HRTEM and electron ptychography
BIO Web of Conferences 129, 04007 (2024)
EMC 2024
https://doi.org/10.1051/bioconf/202412904007
Quantitative comparison of HRTEM and electron
ptychography
Felix Bennemann1, Prof Peter Nellist1, Prof Angus Kirkland1,2
1Department of Materials, University of Oxford, Oxford, United Kingdom,
2Rosalind Franklin Institute, Harwell Science & Innovation Campus, Didcot,
United Kingdom
Background incl. aims
Electron ptychography in the scanning transmission electron microscope
(STEM) has been demonstrated to be capable of providing low-noise phase
images of beam sensitive materials at low dose [1]. For such materials, in
particular biological samples, conventional high-resolution transmission
electron (HRTEM) is the most widely used approach - usually cryo-TEM. The
question then arises of whether ptychography or HRTEM offers the most
dose-efficient imaging approach.
While resolution can be a useful measure for comparing imaging techniques,
it is dependent on the electron dose. For some electron microscopy
techniques, a phase contrast transfer function (PCTF) can be defined to
quantify the technique’s performance with respect to spatial frequency.
However, the PCTF also does not account for the dose used and is not
uniquely defined for common electron ptychography techniques like the
Wigner distribution deconvolution (WDD) method. In this work we introduce
the detective quantum efficiency (DQE), applied to electron microscopy as a
dose independent and sample independent measure of technique
performance.
Historically, the DQE has been used as the ultimate performance
measurement of linear systems [2] ranging from electron detectors to medical
imaging systems. If the incoming noise is pure Poisson noise, it can be
calculated by dividing the signal to noise ratio of the system output squared
(SNRout²) by the signal to noise ratio of the system input squared (SNRin²).
SNRin also represents the signal to noise ratio of an ideal imaging system at
the same dose. In this work the ideal TEM is defined as fully coherent HRTEM
phase contrast imaging with an ideal Zernike phase plate. The SNRout
represents the signal to noise ratio of the various electron microscopy
techniques studied. Even though the signal to noise ratio is dose and sample
dependent, the DQE is not. The DQE can be thought of as the fraction of
incoming quanta contributing to the image.
© The Authors, published by EDP Sciences. This is an open access article distributed under the terms of the Creative Commons
Attribution License 4.0 (https://creativecommons.org/licenses/by/4.0/).
BIO Web of Conferences 129, 04007 (2024)
EMC 2024
https://doi.org/10.1051/bioconf/202412904007
Methods
For the calculation of the empirical DQE, simulations were performed using
the MULTEM package [3]. All 4D-STEM simulations assumed an 80keV beam
energy with a probe step size of 0.15 Å and a semi-angle of convergence angle
α of 22.5 mrad on a 301 by 301 grid of probe positions. A detector size of 128
by 128 pixels was assumed. Reconstructions were performed using the single
sideband (SSB) method, the Wigner distribution deconvolution (WDD)
method and the integrated centre of mass (iCOM) method. As a comparison,
simulations were also performed for high resolution transmission electron
microscopy (HRTEM). The reconstruction methods were evaluated on
coherent and partially coherent datasets containing 500 noise realisations
each at a dose of 4.4M e/Ų. Partial coherence was simulated through the
introduction of a chromatic envelope with Cc = 1.1mm and an energy spread
of 0.4eV leading to a defocus spread of 5.5 nm.
The ground truth was defined as the image from a fully coherent, aberration
free HRTEM phase contrast image. For the purpose of calculating the
empirical DQE of the different methods, a single carbon atom served as a
sample because it provided a continous Fourier transform.
Results
The SSB and WDD electron ptychography methods reach a maximum DQE of
around 23% as shown in the figure. The HRTEM reaches a maximum DQE of
100% at low spatial frequencies. However, through the introduction of a
defocus spread of 5.5 nm, a rapid decay in the DQE of the HRTEM is observed.
The DQE of the HRTEM decays to almost 0% at spatial frequencies above 1α
while the DQE of SSB and WDD remains substantial up to 2α.
It can also be observed that the DQE of both SSB and WDD shows almost no
change through the introduction of the chromatic envelope.
The DQE of iCOM follows the same shape as that of SSB and WDD. However,
across all spatial frequencies it is 3-5% below that of SSB reaching its
maximum at around 18%. It is interesting to note that even with partial
incoherence introduced SSB outperforms fully coherent iCOM.
Conclusion
In this work we have successfully defined a dose and sample independent
framework in which HRTEM and electron ptychography can be compared
against each other using the DQE. We showed that in the absence of partial
incoherence, an HRTEM can achieve a DQE of 100% while SSB and WDD
ptychography have a maximum of around 23%. However, the introduction of
partial coherence shows the weakness of the HRTEM. Considering a defocus
spread of 5.5 nm the DQE of HRTEM remains close to 100% at low spatial
frequencies but decays rapidly at higher spatial frequencies.
2
BIO Web of Conferences 129, 04007 (2024)
EMC 2024
https://doi.org/10.1051/bioconf/202412904007
Graphic:
Keywords:
Ptychography, HRTEM, 4D-STEM
Reference:
[1] Colum M. O’Leary, Gerardo T. Martinez, Emanuela Liberti, Martin J.
Humphry, Angus I. Kirkland, & Peter D. Nellist (2021). Contrast transfer and
noise considerations in focused-probe electron ptychography.
Ultramicroscopy, 221, 113189.
[2] R. C. Jones. Advances in Electronics and Electron Physics XI (Academic
Press. Inc., New York, 1959, p. 121.)
[3] I. Lobato, & D. Van Dyck (2015). Ultramicroscopy, 156, 9-17.
3
(...truncated)