Physiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
ERAY KULCU
MS
1
JANET A. TAMADA
PHD
1
GERARD REACH
0
RUSSELL O. POTTS
PHD
1
MATTHEW J. LESHO
PHD
1
0
INSERM U341, Diabetes Department, H otel-Dieu Hospital
,
Paris, France. City, CA 94063
1
Cygnus
,
Redwood City, California
OBJECTIVE - This study investigated whether glucose readings from a sensor sampling in interstitial fluid differ substantially from blood glucose (BG) values measured at the same time. RESEARCH DESIGN AND METHODS - We have evaluated the relationship between BG and glucose extracted from interstitial fluid using the GlucoWatch (Cygnus, Redwood City, CA) biographer, a device that collects glucose from subcutaneous interstitial space through intact skin by application of a low electric current. We evaluated the relative change in the interstitial glucose (IG) signal (IGS) as measured by the biographer versus BG using a normalized two-point sensitivity index (NSI). RESULTS - The results show that biographer measures of IG differ in time and magnitude from the corresponding BG values. In particular, the biographer values were shifted in time due to instrumental and physiological lag. Results show an average total lag of 17.2 7.2 min for all subjects evaluated. The instrumental lag was 13.5 min, suggesting that physiological lag is 5 min. In addition, when glucose was increasing, the change in IGS was less than that in BG, while when BG was decreasing, the change in IGS was greater than that in BG. CONCLUSIONS - Similar results have been reported for other measures of IG, suggesting that differences reflect physiological variation in glucose uptake, utilization, and elimination in blood and interstitial space. This further evidence of the difference between IG and BG should be considered when interpreting glucose measurements from devices that sample interstitial fluid.
-
G (IG) may differ substantially from when IG was measured by a microdialysis
lucose measured in interstitial fluid ferences. Similar results were obtained
blood glucose (BG) values mea- technique and compared with BG (2).
sured at the same time. Reach et al. (1) The purpose of this study was to
inevaluated dynamic changes in IG in rats vestigate whether these differences
obusing an implanted glucose sensor. Their served in interstitial fluid glucose sensors
results showed that IG was lower than BG can also be observed using an
iontowhen glucose was increasing, yet IG and phoretic, transdermal glucose sensor. We
BG were similar when glucose was declin- have evaluated the relationship between
ing. They developed a physiological BG and glucose extracted from interstitial
push-pull model to describe these dif- fluid using a GlucoWatch (Cygnus,
Redwood City, CA) biographer. This is
an Food and Drug Administration
approved device used to detect trends
and track patterns in glucose levels (35).
It works through a process called reverse
iontophoresis (6), which allows the
biographer to collect glucose samples from
subcutaneous interstitial space through
intact skin by application of a low-level
electric current. When current is applied,
glucose molecules are pulled through the
skin, along with charged molecules
(positive and negative ions) and the
surrounding medium (water). The ions migrate to
gel collection discs placed at the anode
() and cathode () in a single-use
AutoSensor. The glucose molecules are then
collected in these discs for analysis. The
gel collection discs contain the enzyme
glucose oxidase. As glucose enters the
discs, it reacts with the glucose oxidase in
the gel and forms hydrogen peroxide. A
biosensor in contact with each gel
collection disc detects the hydrogen peroxide
generating a current. This current is
integrated, producing a signal in units of
electric charge, nanoCoulombs (nC). The
biographer uses a calibration value
entered by the patient to convert the signal
into a glucose measurement. The
singlepoint calibration is performed with a
traditional BG meter after a 3-h warm-up
period.
RESEARCH DESIGN AND
METHODS Data from existing
clinical studies were used for this analysis (5).
The studies consisted of diabetic (type 1
and insulin-treated type 2) subjects, each
wearing one or two biographers.
Informed consent was obtained from all
subjects. A total of 59 wear periods were
chosen solely on the basis of significant
BG excursion throughout the day and
sufficient available biographer data to
perform time lag analyses. The data were
from studies in a controlled clinical
setting (28 wear periods) and a simulated
home environment (31 wear periods).
The investigator induced mild hyper- and
hypoglycemia during the studies
(between 40 and 450 mg/dl).
The 59 biographer wear periods were
from 51 unique subjects. There were
eight subjects who provided data from
two wear periods. There were 22 men and
29 women, and 32 subjects with type 1
diabetes and 19 with type 2 diabetes. The
means SD of other descriptive statistics
were as follows: age 52 12 years, HbA1c
(prestudy) 7.7 1.6%, and BMI 28.6
4.6 kg/m2.
Time lag
The effect of time lag can potentially
confound the assessment of interstitial and
BG differences. However, when frequent
BG and biographer data are available, it is
possible to calculate the time lag between
the two measurements and eliminate it.
To estimate time lag, IG values were
linearly interpolated to yield values every
minute. BG readings were taken twice per
hour using a HemoCue (Aktiebolaget
Leo, Helsingborg, Sweden) analyzer. The
time shift required to match the
biographer and BG values was then determined
using a method similar to
cross-correlation analysis, where the biographer
glucose data were shifted in time with respect
to the BG data until maximal overlap
was obtained. This analysis yielded an
average total lag of 17.2 7.2 min for the
biographer glucose relative to the BG
value. A similar value was obtained
previously (2).
The total lag is a sum of the
instruFigure 3 A plot of NSI versus BG1 for decreasing glucose values. The regression line (NSI
1.3821 0.0005 BG1, r 0.023, n 489) is also shown.
mental and physiological lag. The
instrumental lag arises from biographers
measurement method, which produces a
glucose value every 20 min by averaging
two 10-min values. Each 10-min period
consists of 3 min of glucose collection and
7 min of glucose detection. Thus glucose
is collected from 20 to 17 min and
10 to 7 min, relative to the biographer
value. This leads to an average
instrumental lag of 13.5 min. The physiological
lag represents the time required for BG to
diffuse from capillaries into the interstitial
space. The physiological lag associated
with the biographer, therefore, is the
difference of total lag and instrumental lag.
In this case, it was 3.7 min (2), which is in
agreement with previously published
estimates of time lag between BG and IG (8).
For the following analysis, this value was
approximated as 5 min. (This analysis
used 1-min interpolation of values, so this
approximation i (...truncated)