Oxygenation-sensitive cardiovascular magnetic resonance (CMR) is a non-contrast technique that allows the non-invasive assessment of myocardial oxygenation. It capitalizes on the fact that deoxygenated hemoglobin in blood can act as an intrinsic contrast agent, changing proton signals in a fashion that can be imaged to reflect the level of blood oxygenation. Increases in O2 saturation increase the BOLD imaging signal (T2 or T2*), whereas decreases diminish it. This review presents the basic concepts and limitations of the BOLD technique, and summarizes the preclinical and clinical studies in the assessment of myocardial oxygenation with a focus on recent advances. Finally, it provides future directions and a brief look at emerging techniques of this evolving CMR field.
Cardiovascular magnetic resonance; Blood-oxygen level-dependent; Microcirculation; Ischemia; Oxygenation
-
Review
Contrast Generation in Oxygenation-Sensitive CMR
It was Linus Pauling who first described an effect of the
oxygenation state of hemoglobin on its magnetic properties
[1]. With respect to its behavior during magnetic resonance
experiments, the de-oxygenation of hemoglobin causes the
molecule to act as an intrinsic paramagnetic contrast
agent resulting in pronounced spin-spin interaction. This
accelerates the decay of transverse magnetization and thus
shortens the spin-spin relaxation time. The term T2*
describes the rate of loss of the transverse signal when
using sequences without refocusing radiofrequency pulses.
Regional iron deposition, hemoglobin degradation products
in tissue hemorrhage or deoxygenated hemoglobin
accelerate this process by their paramagnetic properties, a
phenomenon called the BOLD
(Blood-Oxygen-LevelDependent) effect. This phenomenon can be exploited to
perform BOLD-sensitive or, depending on the context,
oxygenation-sensitive imaging with the aim to detect
changes in tissue oxygenation.
In 1990, Ogawa et al. demonstrated that
oxygenationsensitive MR can be used to detect consequences of very
small blood flow changes in the brain resulting from
external stimuli [2]. In humans, typically several averages
and color-coded maps are used to visualize these changes
[3], a technique which is widely known as functional brain
MR. Accordingly, myocardial deoxygenation or ischemia is
characterized by a net relative increase of de-oxygenated
hemoglobin in the capillary blood and thus leads to T2*
shortening, which can be visualized by T2* maps or
by a regional signal loss in T2*-weighted MR images
(i.e. images acquired by protocols sensitive to decreased
regional field homogeneity). Conversely, a decrease of the
proportion of de-oxygenated hemoglobin (for example by
inducing vasodilation without a matching increase of
myocardial oxygen demand) causes a relative decrease of
de-oxygenated hemoglobin and leads to an increase of
T2* in this territory and hence to an increased signal
intensity in oxygenation-sensitive images.
It is important to keep in mind that the observed changes
reflect changes in (mainly the venous compartment of the)
capillary bed and therefore strictly do not represent the
actual cell (i.e. cardiomyocytes). Yet the oxygenation of
the capillary blood directly reflects the balance of oxygen
supply and demand and therefore can be understood as a
direct marker of tissue oxygenation.
Technical aspects of oxygenation-sensitive CMR
While image quality is frequently affected in echo-planar
sequences [4] and other, T2-weighted sequences [5] by
motion or susceptibility artifacts (especially along the
lung-heart interface), steady-state-free-precession sequences
have shown promising results, allowing for a simultaneous
acquisition of morphological, functional and
oxygenationsensitive data of the heart [6]. Of note, the
oxygenationsensitive contrast in images acquired by this sequence is
directly dependent on the repetition time TR [7]. With an
appropriate TR, the obtained image quality is much more
consistent when compared to other BOLD-sensitive
sequences, though at the expense of reduced sensitivity to
the BOLD effect and thus lower oxygenation contrast.
Further studies may have to fine-tune the balance between
sensitivity to oxygenation changes and susceptibility
artifacts.
Importantly, the magnitude of the BOLD effect (i.e. the
measurable effect size) largely depends on the strength of
the static magnetic field. While the overall susceptibility to
field inhomogeneity at higher field strengths may cause
artifacts, it also accounts for a higher sensitivity to changes
induced by paramagnetic effects. Higher field strengths
thus improve the sensitivity of MR to the BOLD effect.
Data by Dharmakumar et al. indicate that the sensitivity
to detect changes in myocardial oxygenation may increase
by a factor of about 2.5 when moving from 1.5T to 3T [8].
With respect to a clinical application in patients with
suspected myocardial ischemia, it is very important to be
aware of the significant limitations of current diagnostic
techniques to verify the hemodynamic relevance of coronary
artery disease. Because of a lack of diagnostic targets on
the (cellular) level of actual ischemia, imaging techniques
use surrogate markers such as stress-inducible dysfunction
(echocardiography), changes of blood inflow characteristics
(echocardiography, first-pass perfusion CMR) or metabolic
changes (nuclear cardiology techniques). All these are
surrogates and thus cannot directly reflect an ischemic
response of myocardial tissue, while oxygenation-sensitive
CMR offers exactly that.
Experimental and pre-clinical studies
In 1993, Atalay et al. used a Langendorff heart
model to show that the signal intensity in images
sensitive to the BOLD effect (in that paper called
susceptibility-dependent) is closely correlated to blood
oxygenation [9]. Wendland et al. induced hypoxia in
rats and demonstrated changes in the blood and the
myocardium caused by the BOLD effect [10].
In July 1996, Niemi et al. observed in healthy volunteers,
that the increase of myocardial blood flow as induced by
the infusion of the vasodilator dipyridamole was closely
correlated with a myocardial signal intensity increase in
images obtained by a T2*-sensitive single-shot echo planar
imaging (EPI) sequence [11]. Yet, another study, published
in the same journal issue by (...truncated)