Transformative effects of higher magnetic field in Fourier transform ion cyclotron resonance mass spectrometry
N. Murat Karabacak
0
1
4
Michael L. Easterling
0
1
2
Nathalie Y. R. Agar
0
1
3
Jeffrey N. Agar
0
1
4
0
Address reprint requests to Dr. J. N. Agar,
Department of Chemistry, Brandeis University
,
MS015
, 415 South Street, Waltham,
MA 02454, USA
1
Published online March 31, 2010 Received November 16, 2009 Revised March 4, 2010 Accepted March 4, 2010
2
Bruker Daltonics, Billerica,
MA, USA
3
Department of Neurosurgery, Brigham and Women's Hospital, Harvard Medical School
,
Boston, MA, USA
4
Department of Chemistry and Volen Center for Complex Systems, Brandeis University
, Waltham,
MA, USA
The relationship of magnetic field strength and Fourier transform ion cyclotron resonance mass spectrometry performance was tested using three instruments with the same design but different fields of 4.7, 7, and 9.4 tesla. We found that the theoretically predicted transformative effects of magnetic field are indeed observed experimentally. The most striking effects were that mass accuracy demonstrated second to third order improvement with the magnetic field, depending upon the charge state of the analyte, and that peak splitting, which prohibited automated data analysis at 4.7 T, was not observed at 9.4 T. (J Am Soc Mass Spectrom 2010, 21, 1218 -1222) 2010 American Society for Mass Spectrometry
-
important instrument performance metrics can be
affected by factors with no- or convoluted field
dependences. For example, the relationship between mass
accuracy and resolution is blurred by anything that
affects ideal peak shapes [1517].
There are isolated literature examples of what
Marshall terms the transformative effects of higher
magnetic fields. For example, post processing of 14.7 T
FTMS spectrum to simulate a 7 T FTMS spectrum
revealed that three compounds that were identifiable at
14.7 T appeared to be a single compound at 7 T [18]. In
another example, the instruments sensitivity for a
complex and highly charged sample (29 kDa protein
dissociation products) significantly increased when
going from 6 T to 9.4 T FT-ICR [19]. On the other hand,
despite the availability of 15 T instruments, the highest
published resolving powers of 8,000,000 (m/z 1148)
[20], 17,000,000 (m/z 1084.5) [21], 1,000,000 (m/z
12,360) [22] were acquired (in one case 11 years ago [20])
at 9.4 T or 7 T, illustrating the importance of factors
other than magnetic field. While the effects of field
homogeneity upon instrument performance have been
characterized [8], we are aware of no publication that
systematically explores the relationship of FT-ICR MS
performance and gross magnetic field. Here, we make
an empirical determination of the merit of higher
magnetic field, using the same instrument type, the same
acquisition and processing software and methods, the
same methods for optimization, the same user, and
employing magnetic fields from 4.7 to 9.4 T. Moreover,
we perform a meta-analysis of reported mass accuracies
of FTMS instruments ranging from 1 to 15 T, which is
consistent with our experimental findings.
F spectrometry [1] (FT-ICR MS) is currently the
ourier transform ion cyclotron resonance mass
highest resolution mass spectrometry method.
The cyclotron frequency of ions in a given magnetic
field (eq 1) gives a very accurate measure of the mass to
charge ratio of ions.
Numerous performance parameters in FT-ICR MS are
predicted to improve with magnetic field, including
linear improvements in mass resolving power and
acquisition speed, and higher order improvements in
mass accuracy, dynamic range, kinetic energy, and peak
coalescence [27]. All of these parameters combine to
determine the figure of merit of a mass analyzer, but not
in a manner that is easily predicable. Improving FT-ICR
MS performance is critical in many fields such as
proteomics, petroleomics, and MALDI imaging, and in
many cases enables, rather than improves, analytical
capabilities.
In practice, however, a component with no a priori
field dependence, for example field homogeneity [8]
vacuum strength, or acquisition speed, which set
fundamental limits on resolution [2], or a phenomenon
with a convoluted electrical and magnetic field
dependence such as peak coalescence [9] or phase locking [7,
10 14], can become the limiting factor and the
determinant of performance. In addition, the relationship of
All chemicals except ubiquitin (Boston Biochem,
Cambridge, MA, USA) were purchased from Sigma-Aldrich
(St. Louis, MO, USA).
This study involved three separate commercial
hybrid q-FT-ICR mass spectrometers (Bruker Apex-qe)
with the same cell infinity ion trap geometry [23],
electrospray source (Apollo 2), ion transfer optics,
vacuum elements (with readings of below 4 10 10 mbar
in the ICR cell during this study) but with different
magnetic fields. A sodium formate (Sigma) 0.01 mg/mL
solution was prepared in 50% acetonitrile 0.1% formic
acid. A complex peptide and protein mixture with a
concentration ladder spanning four orders of
magnitude was prepared in 50% acetonitrile 0.1% formic acid
and introduced by a syringe pump at 2 L/min. This
mixture consisted of Substance P (1 M), insulin (0.1
M), orexin B (0.01 M), ubiquitin (0.1 M),
angiotensin (0.1 nM), myoglobin (1 M), human Cu/Zn
superoxide dismutase (0.05 M), ribonuclease A (0.5 M) and
lysozyme (0.01 M). One M Substance P was used for
ultra-high resolution narrowband experiment.
Instruments were tuned and operated using our
routine techniques and parameters, which involved the
following steps. Initially, electrospray source and
transfer parameters were optimized for maximum signal
magnitude. Static trapping [24] at relatively low
trapping potentials (1.0 V/0.8 V for 4.7 T, 0.9 V/0.8 V for 7.0
T and 1.2 V/1.2 V for 9.4 T for front/rear
trapping plates) was used. These trapping potentials were
30% greater than potential where no signal was
observed (and presumably little trapping occurred). Next,
excitation amplitude was tuned for each instrument by
determining the excitation amplitude that yielded
maximum mass accuracy when internal calibration was
performed using sodium formate clusters. Resulting
amplitudes were 1.25 dB for 9.4 T, 4.5 dB for 7.0 T and
9.75 dB for 4.7 T. Using these optimized parameters,
spectra of the peptide and protein mixture were
acquired. Ions were directly infused into an electrospray
source at 2 L/min flow rate, externally accumulated in
the source and collision cell hexapole for 1 s each, and
transferred to the ICR cell using ion transfer optics that
had been previously optimized for signal intensity.
Chirp excitation and image charge detection was
performed. Three, 2-M-word datasets in the range of m/z
274 3000 were acquired and 4, 16, or 64 scans were
averaged for each dataset. The FID was multiplied by a
sine bell apodization function and was Fourier
transformed. The narrowband experiment (Figure 3)
involved a 32-k-word dataset covering 3.0 Thomson (Th)
and was averaged 325 times.
We tried to eliminate bias during data analysis in the
following way. Only the p (...truncated)