Single lever Humphrey A.D.E.lowflow universal anaesthetic breathing system
David Humphrey MBBS(Lond.) DA (S.A.)
0
John G. Brock-Utne MA MB BCH (TCD) FFA (S.A.) (Bergen)
0
John W. Downing MB BCH FFA ( S . A . ) FFARCS (Eng.)
0
0
From the Faculty of Medicine, Departments of Anaes- thetics and Physiology, University of Natal
,
PO Box 17039, 4013 Congella, Durban
,
South Africa
,
where Preliminary reports were presented at the 1985 South African Society of Anaesthetists Congress, Durban and at the 1985 18th Scandinavian Society of Anaesthesiol- ogists Congress
,
Iceland
A clinical trial involving ten anaesthetized adultpatients was conducted during controlled ventilation using the Humphrey A.D.E. system in the Mapleson "E" mode (lever down). With each patient acting as his or her own control, the parallel (non-coaxial) and coaxial versions of the single lever Humphrey A.D.E. system were compared, using capnography, to the Bain system (Mapleson D/E). All three systems behaved similarly with predicta. ble patient normocarbia when a fresh gas flow of 70 ml.kg- t.min-t was used. The A.D.E. system has the added advantage that the switch from controlled to spontaneous ventilation (or vice versa) is achieved quickly and simply. Spontaneous, assisted or automatic controlled ventilation could be instituted at any time merely by the appropriate lever position.
-
ABBREVIATIONS
A.D.E. = Humphrey A.D.E. anaesthetic
breath
ing system
_A.D.E. = A.D.E. set in A mode
A._D.E. = A.D.E. set in D mode
A.D._E. = A.D.E. set in E mode
FGF = Fresh gas flow
PICO2 = Minimum inspired carbon dioxide
tension
PI~CO2 = End-expired (or end-tidal) carbon
dioxide tension
PaCOz = Arterial carbon dioxide tension
Humphrey etal.: CONTROLLED VENTILATION WITH THE A.D.E. BREATHING SYSTEM
FIGURE I Photograph of the single lever parallel
(noncoaxial) version of the Humphrey A.D._E.system set in the "E"
mode (lever down). Both the reservoir bag and expiratory
valve are automatically excluded, thus making simple tubes of
the inspiratory and expiratory limbs (see Figures 2-4), While
the ventilator remains attached to the expiratory ventilator
port on the underside at all times, it is included in circuit only
with the lever down, Coaxial tubing can replace the parallel
tubes if desired (Figure 3).
only 51 m l . k g - t . m i n - t to p r e v e n t rebreathing o f
alveolar g a s e s , a n d to be m o r e efficient t h a n the
Magill s y s t e m w h i c h required a m e a n flow o f
71 m l . k g - t . m i n - I . T h e Bain s y s t e m w a s m a r k e d l y
less efficient than b o t h A . D , E . or Magill s y s t e m s ,
and required a m e a n F G F o f 150 m l - k g - l . m i n - I .
However, d u r i n g controlled ventilation, similar
function o f the A . D . E, a n d t h e B a i n s y s t e m s c a n be
predicted, since both a r e c o n v e r t e d into M a p l e s o n E
s y s t e m s * that are functionally two s i m p l e t u b e s
(Figure 4). T h e B a i n s y s t e m h a s b e e n u s e d
extensively in this m o d e b e c a u s e PaCO2 can b e
predictably determined by the fresh g a s flow set b y the
anaesthetist. Identical p e r f o r m a n c e o f the original
dual lever A . D . E , s y s t e m a n d the B a i n s y s t e m h a s
been confirmed d u r i n g controlled ventilation. 6 W e
now present the results o f c a p n o g r a p h s t u d i e s o n
anaesthetised adult patients d u r i n g controlled
ventilation with both v e r s i o n s o f the n e w single lever
*See appendix.
FIGURE 2 Diagram of the parallel(non-coaxial) single lever Humphrey A.D.E. system with an exploded view of the eytinder
mechanism. (a) When the lever is upright (A mode, spontaneous and manual respiration) and inspiratcyryreservoir bag and
expiratoryvalve are in circuit, while the ventilator is excluded. (b) When the lever is down (E mode), the ventilator is now included
with the exclusion of the reservoir bag and exhaus! valve. The system is simply two tubes, as is the Bain circuit during
controlled ventilation (see Figure 4). Note that the pressure-limiting device stays in circuit in either mode. FGF = fresh gas flow.
(Reproduced from Humphrey,4 with permission.
FIGURE 3 Diagram of the single lever coaxial Humphrey A_.D.E. system. The same main body as in Figure 2 is used but the
parallel tubing is replaced by a coaxial set. The position of the lever and function of the system remain identical.
FIGURE 4 Diagrams of the (a) Bain, (b) coaxial A.D.E_. and
(c) non-coaxial parallel A.D._E. systems showing the
func!ional parts when used for controlled ventilation. All are simply
two tubes (Mapleson E systems), since the valve on each
system has been closed (Bain) or excluded (A.D._E.) and the
reservoir bag removed (Bain) or excluded (A.D._E.).
Methods
A similar protocol and anaesthetic induction
technique as has been previously presenteds was
employed. Ten patients scheduled for elective
peripheral surgery and of ASA physical status I were
selected. Each patient gave informed consent.
Anaesthesia was induced with thiopentone, the
trachea was intubated and anaesthesia was
maintained with halothane and nitrous oxide with
oxygen. However, in this study, the use of narcotic
analgesics during anaesthesia was permitted as
required. Muscle relaxation was achieved with
alcuronium 0.15 to 0.25 mg'kg - t IV as an initial
dose, with smaller increments being given as
necessary.
Ventilation was controlled using a fluid logic,
time cycled, flow generator Nuffield 200 ventilator
(Penlon Ltd., Abingdon, Oxon, England) attached
to the expiratory valve-bypass outlet of the A.D._E.
system via two standard 22 mm corrugated tubes
with a total volume of about one litre. This
ventilator is of the "bag squeezer" variety and is suitable
Humphreyetal.: C O N T R O L L E D V E N T I L A T I O N W I T H T H E A , D . E . B R E A T H I N G SYSTEM
for Bain type and circle systems. Tidal volumes
were calculated as normally recommended for the
Bain system at approximately 10ml.kg-l-min -~
with the respiratory rate set between 12 and 15
breaths.min - l , and fresh gas flows adjusted to
70ml.kg-J'min - j . Both A.D._E. systems and the
Bain circuit were compared twice in each patient,
the initial order of testing being randomized, six
patients being connected to an A.D._E. system first.
Clinical trials
Investigations were conducted with both parallel
and coaxial versions of the single lever A.D.E_.
systems set in their Mapleson E mode with the lever
in the down position (Figures 1-3). With the
ventilator connected to the expiratory valve-bypass
outlet, tidal volumes, ventilation frequency and
fresh gas flows were set, as above. The Bain system
was used in the conventional way with the ventilator
attached to the expiratory Iimb with the reservoir
bag removed and the exhaust valve fully closed.
Within patient comparisons in ten subjects were
made between both A.D.E_. systems and the Bain,
without alteration of any of the ventilation or fresh
gas parameters.
Both versions of the single lever A.D.E. systems
were also examined towards t (...truncated)